BIS™ Advance Monitor
WARNING: Whenever an event such as spillage of blood or solutions occurs, re-
test ground leakage current before further use.
WARNING: Power supply is internally fused. Replace power supply only with
the power supply designated for this monitor, Medtronic PN PMB4000PWS.
WARNING: When connecting external equipment (e.g., data capture
computer), the system leakage current must be checked and must be less than
the IEC 60601-1-1 limit. Only approved medical devices which comply to
60601-1 (or USB flash drives) shall be connected to the monitor or docking
station.
WARNING: Using accessories other than those specified may result in increased
electromagnetic emissions or decreased electromagnetic immunity of the BIS™
Advance monitor.
WARNING: The use of accessory equipment not complying with the equivalent
safety requirements of this equipment may lead to a reduced level of safety of
the resulting system. Consideration relating to the choice shall include:
oUse of the accessory in the patient vicinity.
Evidence that the safety certification of the accessory has been performed in
accordance to the appropriate IEC 60601-1 and/or IEC 60601-1-1 harmonized
national standard.
Caution: Keep power cord, plug and socket clear in case an urgent power
supply disconnection is required.
Caution: When connecting or disconnecting the BISx module, take care not to
touch the exposed contacts of either connector. Damage due to electrostatic
discharge may result.
1.2.3. Battery Cautions
WARNING: Charge the monitor fully before your first use of the monitor, by
attaching the monitor to AC power with the battery installed. See 3.3 Battery
Operation on page 60.
WARNING: The monitor should always be operated with the battery installed
in order to provide back-up power in the event of a momentary or temporary
power outage.
WARNING: Improper operation may cause damage to the battery or endanger
the user.