Operator's Manual
Table Of Contents
- 1. Introduction
- 1.1. Introduction
- 1.2. Safety Information
- 1.3. Device Description
- 1.4. Intended Purpose
- 1.5. BIS™ Advance Monitor System Indications and Contraindications
- 1.6. Patient Target Groups
- 1.7. Intended Users
- 1.8. Essential Performance
- 1.9. Adverse Events (Residual Risks)
- 1.10. The BIS™ Advance Parameters
- 1.11. Symbols
- 2. The BIS™ Advance Monitor — Equipment and Supplies
- 2.1. The BIS™ Advance System
- 2.2. Parts of the System
- 3. Installation and Preparation for Use
- 3.1. Operating Environment
- 3.2. Power Requirements and Battery Use
- 3.3. Battery Operation
- 3.4. Mounting the Docking Station
- 3.5. Preparing the Docking Station
- 3.6. Preparing the Monitor
- 3.7. Home Screen
- 3.8. Four-Channel Monitoring
- 3.9. Home Screen Trend Graphs
- 3.10. Home Screen Numeric Section
- 3.11. Home Screen Menus
- 3.12. Home Screen Options
- 3.13. Secondary Trend Selection
- 3.14. Home Screen Audio Alarm Settings
- 3.15. Settings and Maintenance
- 3.16. Info Options
- 3.17. Quick Reference Checklist
- 4. Operating the BIS™ Advance Monitoring System
- 4.1. Preparing for Operation
- 4.2. Cases
- 4.3. Alarms and Messages
- 4.4. Chart Data
- 4.5. Menu Map
- 4.6. Institutional Settings
- 5. Data Storage, Transfer, and Export
- 5.1. Cybersecurity and Data Integrity
- 5.2. Monitor Data Memory
- 5.3. BISx Data Memory
- 5.4. Types of Exported Data
- 5.5. Live Case Export
- 5.6. Saved Cases Export
- 5.7. Recording a Snapshot Event
- 5.8. BIS™ Saved Data Files
- 5.9. Logs Export
- 5.10. Viewing and Printing Saved Data in PDF Format
- 6. Service and Maintenance
- 6.1. Introduction
- 6.2. Servicing the BIS™ Advance Monitor
- 6.3. Battery Handling
- 6.4. Cleaning the BIS™ Advance Monitor
- 6.5. Instrument Identification
- 7. Diagnostics and Troubleshooting
- 7.1. Diagnostics
- 7.2. Troubleshooting
- 7.3. Administrator Mode
- 7.4. Demo Mode
- 7.5. Glossary
- 7.6. Specifications
- 7.6.1. General Specifications
- 7.6.2. Operating Environments
- 7.6.3. Power Supply
- 7.6.4. Battery Specifications
- 7.6.5. Controls and Connectors
- 7.6.6. Alarms
- 7.6.7. Display
- 7.6.8. Sound Pressure Data
- 7.6.9. BIS Specifications
- 7.6.10. EEG Specifications
- 7.6.11. BISx Module Specifications
- 7.6.12. Software/GUI Specifications
- 7.7. Electromagnetic Compatibility Specifications
- 7.8. Product Compliance
- 7.9. BIS™ Advance Components, Accessories, and Documentation
- 7.10. Warranty
- 7.11. Software License Agreement
Alarms and Messages
BIS™ Advance Monitor
151
Alarm Message Possible Causes Corrective Actions
Alarm
Priority
PRESS CENTER
OF ELECTRODES
{0} TO ENABLE
MONITORING
One (or more) of
the sensor
electrodes is not in
contact. The
relevant electrode
number will
appear instead of
the parentheses.
Press the center of the
electrode listed on the BIS
sensor to initiate contact
and enable monitoring.
LOW
TEMPORARILY
UNABLE TO
MEASURE
ELECTRODES {0}
The system is
temporarily
unable to measure
data from one of
the electrodes on
the sensor. The
relevant electrode
number will
appear instead of
the parentheses.
If the issue does not correct
itself, replace the sensor.
LOW
POSSIBLE PIC
PROBLEM
Replace PIC or
check cable
using sensor
simulator
Ground electrode
impedance is high
and impedance of
other electrodes is
low.
• Defective PIC.
1. Test another PIC with
sensor simulator.
2. If failure persists, replace
simulator and retest.
3. If failure persists, replace
PIC.
LOW
POSSIBLE PIC
PROBLEM
Replace PIC
One of the
electrodes (other
than ground) has
high impedance.
• Defective PIC.
• Defective sensor
simulator.
1. Test another PIC with
sensor simulator.
2. If failure persists, replace
simulator and retest.
3. If failure persists, replace
PIC.
LOW