Operator's Manual
Table Of Contents
- 1. Introduction
- 1.1. Introduction
- 1.2. Safety Information
- 1.3. Device Description
- 1.4. Intended Purpose
- 1.5. BIS™ Advance Monitor System Indications and Contraindications
- 1.6. Patient Target Groups
- 1.7. Intended Users
- 1.8. Essential Performance
- 1.9. Adverse Events (Residual Risks)
- 1.10. The BIS™ Advance Parameters
- 1.11. Symbols
- 2. The BIS™ Advance Monitor — Equipment and Supplies
- 2.1. The BIS™ Advance System
- 2.2. Parts of the System
- 3. Installation and Preparation for Use
- 3.1. Operating Environment
- 3.2. Power Requirements and Battery Use
- 3.3. Battery Operation
- 3.4. Mounting the Docking Station
- 3.5. Preparing the Docking Station
- 3.6. Preparing the Monitor
- 3.7. Home Screen
- 3.8. Four-Channel Monitoring
- 3.9. Home Screen Trend Graphs
- 3.10. Home Screen Numeric Section
- 3.11. Home Screen Menus
- 3.12. Home Screen Options
- 3.13. Secondary Trend Selection
- 3.14. Home Screen Audio Alarm Settings
- 3.15. Settings and Maintenance
- 3.16. Info Options
- 3.17. Quick Reference Checklist
- 4. Operating the BIS™ Advance Monitoring System
- 4.1. Preparing for Operation
- 4.2. Cases
- 4.3. Alarms and Messages
- 4.4. Chart Data
- 4.5. Menu Map
- 4.6. Institutional Settings
- 5. Data Storage, Transfer, and Export
- 5.1. Cybersecurity and Data Integrity
- 5.2. Monitor Data Memory
- 5.3. BISx Data Memory
- 5.4. Types of Exported Data
- 5.5. Live Case Export
- 5.6. Saved Cases Export
- 5.7. Recording a Snapshot Event
- 5.8. BIS™ Saved Data Files
- 5.9. Logs Export
- 5.10. Viewing and Printing Saved Data in PDF Format
- 6. Service and Maintenance
- 6.1. Introduction
- 6.2. Servicing the BIS™ Advance Monitor
- 6.3. Battery Handling
- 6.4. Cleaning the BIS™ Advance Monitor
- 6.5. Instrument Identification
- 7. Diagnostics and Troubleshooting
- 7.1. Diagnostics
- 7.2. Troubleshooting
- 7.3. Administrator Mode
- 7.4. Demo Mode
- 7.5. Glossary
- 7.6. Specifications
- 7.6.1. General Specifications
- 7.6.2. Operating Environments
- 7.6.3. Power Supply
- 7.6.4. Battery Specifications
- 7.6.5. Controls and Connectors
- 7.6.6. Alarms
- 7.6.7. Display
- 7.6.8. Sound Pressure Data
- 7.6.9. BIS Specifications
- 7.6.10. EEG Specifications
- 7.6.11. BISx Module Specifications
- 7.6.12. Software/GUI Specifications
- 7.7. Electromagnetic Compatibility Specifications
- 7.8. Product Compliance
- 7.9. BIS™ Advance Components, Accessories, and Documentation
- 7.10. Warranty
- 7.11. Software License Agreement
Alarms and Messages
150
BIS™ Advance Monitor
Alarm Message Possible Causes Corrective Actions
Alarm
Priority
BATTERY LOW
Connect to
power supply
There is only a
short amount of
battery usage time
remaining.
Restore AC power to avoid
automatic shutdown.
LOW
DSC SELF-TEST
FAILED
Replace BISx
The DSC Self-Test
process failed.
Replace the BISx module. LOW
BISx
INITIALIZATION
ERROR
Disconnect and
reconnect BISx
o Poor connection
between BISx
module monitor
cable (MIC) and
monitor.
o Defective BISx.
o Defective
monitor.
1. Disconnect and
reconnect the adapter
cable. Unplug and connect
the adapter cable.
2. Follow on-screen
instructions. If necessary,
power OFF and unplug
power cord to shut down
monitor completely. Then
plug in and restart monitor.
3. Replace the BISx.
4. Replace the monitor.
LOW
UNABLE TO
READ SENSOR
Replace the
sensor
The monitor is
unable to read
data from the BIS™
sensor.
Replace the sensor. LOW
UNSUPPORTED
SENSOR TYPE
Replace the
sensor
The sensor is not
compatible with
the monitor
configuration.
Replace sensor. Reference
Table 33. BIS™ Advance
Monitor Components and
Sensors on page 252 for a
list of approved sensors.
Ensure that the sensor used
is compatible with the
monitor configuration.
LOW