Operator's Manual
Table Of Contents
- 1. Introduction
- 1.1. Introduction
- 1.2. Safety Information
- 1.3. Device Description
- 1.4. Intended Purpose
- 1.5. BIS™ Advance Monitor System Indications and Contraindications
- 1.6. Patient Target Groups
- 1.7. Intended Users
- 1.8. Essential Performance
- 1.9. Adverse Events (Residual Risks)
- 1.10. The BIS™ Advance Parameters
- 1.11. Symbols
- 2. The BIS™ Advance Monitor — Equipment and Supplies
- 2.1. The BIS™ Advance System
- 2.2. Parts of the System
- 3. Installation and Preparation for Use
- 3.1. Operating Environment
- 3.2. Power Requirements and Battery Use
- 3.3. Battery Operation
- 3.4. Mounting the Docking Station
- 3.5. Preparing the Docking Station
- 3.6. Preparing the Monitor
- 3.7. Home Screen
- 3.8. Four-Channel Monitoring
- 3.9. Home Screen Trend Graphs
- 3.10. Home Screen Numeric Section
- 3.11. Home Screen Menus
- 3.12. Home Screen Options
- 3.13. Secondary Trend Selection
- 3.14. Home Screen Audio Alarm Settings
- 3.15. Settings and Maintenance
- 3.16. Info Options
- 3.17. Quick Reference Checklist
- 4. Operating the BIS™ Advance Monitoring System
- 4.1. Preparing for Operation
- 4.2. Cases
- 4.3. Alarms and Messages
- 4.4. Chart Data
- 4.5. Menu Map
- 4.6. Institutional Settings
- 5. Data Storage, Transfer, and Export
- 5.1. Cybersecurity and Data Integrity
- 5.2. Monitor Data Memory
- 5.3. BISx Data Memory
- 5.4. Types of Exported Data
- 5.5. Live Case Export
- 5.6. Saved Cases Export
- 5.7. Recording a Snapshot Event
- 5.8. BIS™ Saved Data Files
- 5.9. Logs Export
- 5.10. Viewing and Printing Saved Data in PDF Format
- 6. Service and Maintenance
- 6.1. Introduction
- 6.2. Servicing the BIS™ Advance Monitor
- 6.3. Battery Handling
- 6.4. Cleaning the BIS™ Advance Monitor
- 6.5. Instrument Identification
- 7. Diagnostics and Troubleshooting
- 7.1. Diagnostics
- 7.2. Troubleshooting
- 7.3. Administrator Mode
- 7.4. Demo Mode
- 7.5. Glossary
- 7.6. Specifications
- 7.6.1. General Specifications
- 7.6.2. Operating Environments
- 7.6.3. Power Supply
- 7.6.4. Battery Specifications
- 7.6.5. Controls and Connectors
- 7.6.6. Alarms
- 7.6.7. Display
- 7.6.8. Sound Pressure Data
- 7.6.9. BIS Specifications
- 7.6.10. EEG Specifications
- 7.6.11. BISx Module Specifications
- 7.6.12. Software/GUI Specifications
- 7.7. Electromagnetic Compatibility Specifications
- 7.8. Product Compliance
- 7.9. BIS™ Advance Components, Accessories, and Documentation
- 7.10. Warranty
- 7.11. Software License Agreement
Alarms and Messages
BIS™ Advance Monitor
147
Alarm Message Possible Causes Corrective Actions
Alarm
Priority
SENSOR
GROUND FAULT
Problem is
detected relating
to sensor ground
element. A
continuous
ground fault has
occurred for
longer than 8
seconds.
1. Disconnect and examine
sensor connection. Clean
any contamination present.
2. Replace sensor if
necessary.
3. Replace PIC.
4. Replace BISx.
HIGH
MONITOR
TEMPERATURE
LIMIT EXCEEDED
If the problem
persists, consider
replacing
monitor
Monitor
temperature limit
has been
exceeded.
If the problem persists, the
monitor may need to be
replaced. Contact technical
support.
MEDIUM
BATTERY IS
CRITICALLY LOW
Monitor may
shut down at any
time
There are only a
few minutes of
battery usage time
left. Monitor may
shut down at any
time.
Restore AC power to avoid
automatic shutdown and
continue working.
MEDIUM
SR ALARM
SR above defined
threshold
SR value is above
the defined
threshold.
Check patient status. LOW
SEF ALARM
SEF above
defined range
SEF value is above
the defined range.
Check patient status. LOW
SEF ALARM
SEF below
defined range
SEF value is below
the defined range.
Check patient status. LOW