BIS™ Advance Monitor
the user resets the monitor to Institutional Defaults (using the RESET SETTINGS
button as described in 4.6 Institutional Settings on page 180) or an administrator
resets the monitor to Factory Defaults (using the RESET SETTINGS button as
described in 4.6 Institutional Settings on page 180).
To enter the screen, click Main Menu>Settings and Maintenance>
Settings>Advanced>Impedance. To change the status of this option, slide the
switch left to turn OFF (switch will then appear grey) or to the right to turn ON
(switch will then appear green) and click APPLY.
This option is available only when a sensor is attached to the monitor.
Sensor Check Values
The Sensor Check Values screen will permit the user to turn the display of sensor
check values on the sensor check screen off or on. By default, display of the values is
off.
To enter the screen, click Main Menu>Settings and
Maintenance>Settings>Advanced>Sensor Check Values. To change the
status of this option, slide the switch left to turn OFF (switch will then appear grey) or
to the right to turn ON (switch will then appear green) and click APPLY.
This option is available whether or not a sensor is connected to the monitor.
Display Settings
The Display Settings screen will permit the user to select a particular trend display
layout. By default, the layout for a 2-channel system is a 2-trend layout with BIS and
EEG. For a 4-channel system, the default layout is a 3-trend layout with BIS, ASYM and
EEG shown. The defaults can be adjusted in the Administrator Mode; reference 4.6
Institutional Settings on page 180.
To enter the screen, click Main Menu>Settings and Maintenance>
Settings>Advanced>Display Settings. To change the layout, click Change. This
will open the Display Settings screen (also accessible by clicking Main
Menu>Display Settings). To change the display , select a layout from the bar at
left, and click Apply.
For a description of available layouts, reference 3.12 Home Screen Options on page
121.