132
BIS™ Advance Monitor
Trends graphical display, make the desired changes on this screen and click APPLY
to apply the changes.
The EEG Trends Graphical Display may be changed only when a sensor is attached to
the monitor and the current display settings include an EEG waveform display.
Parameter Options Default
EEG Amplitude +/- 50 [µV]
+/- 100 [µV] (only when full
screen EEG layout is selected
as described in 3.12 Home
Screen Options on page 121)
+/- 50 [µV]
EEG Scale See Table 10. EEG Display Scale
Options on page 132
See Table 10. EEG Display
Scale Options on page 132
Number (#) of
Channels
1,2,4 Depends on EEG scale
and selection of 2-
channel or 4-channel
monitoring
EEG Filters On/Off, using slide button to
toggle
On
The software shall allow the user to choose the EEG display scale per number of
channels, according to the details seen in Table 10. EEG Display Scale Options, below.
Table 10. EEG Display Scale Options
Display Scale Full Screen (1:1) Half Screen (1:2)
One Third
Screen (1:3)
2 Channel Display Scale
2 µV/mm 1 (no. of channels) 1 NA
5 µV/mm 1,2 1,2 1