BIS™ Advance Monitor
Alarm Pause (Temporary
Alarm Silence)
Pauses the audio alarm
for default time of 120
seconds, with
countdown
Alarm Off (Permanent Alarm
Silence)
Turns off the audio
alarm, until it is turned
on again or device is
restarted
Alarm Volume adjustment
Opens window in which
alarm volume can be
adjusted
For more information about alarm settings, reference 4.3 Alarms on page 143.
3.15. Settings and
Maintenance
The monitor provides a number of settings adjustments that can be performed in the
Settings and Maintenance screen by the user. Please note that these changes will be
in force only for the current case or until restart of the monitor (whichever comes
first). Once the monitor is shut down and restarted or a new case is started, the
monitor will default to institutional Defaults.
To change institutional Settings on a more permanent basis, reference 7.3
Administrator Mode on page 218.
WARNING: The software will reset all settings with defined default values to
Institutional Defaults upon the end of each case, excluding Institutional
Default Language settings. The software will also reset all settings (including
language) to Institutional Defaults upon the restart of the monitor. However, if
a sensor for which certain settings were set (as described in 3.15.3 Advanced
User Settings on page 131, and whether changes were performed by a user or