Operator's Manual
Table Of Contents
- 1. Introduction
- 1.1. Introduction
- 1.2. Safety Information
- 1.3. Device Description
- 1.4. Intended Purpose
- 1.5. BIS™ Advance Monitor System Indications and Contraindications
- 1.6. Patient Target Groups
- 1.7. Intended Users
- 1.8. Essential Performance
- 1.9. Adverse Events (Residual Risks)
- 1.10. The BIS™ Advance Parameters
- 1.11. Symbols
- 2. The BIS™ Advance Monitor — Equipment and Supplies
- 2.1. The BIS™ Advance System
- 2.2. Parts of the System
- 3. Installation and Preparation for Use
- 3.1. Operating Environment
- 3.2. Power Requirements and Battery Use
- 3.3. Battery Operation
- 3.4. Mounting the Docking Station
- 3.5. Preparing the Docking Station
- 3.6. Preparing the Monitor
- 3.7. Home Screen
- 3.8. Four-Channel Monitoring
- 3.9. Home Screen Trend Graphs
- 3.10. Home Screen Numeric Section
- 3.11. Home Screen Menus
- 3.12. Home Screen Options
- 3.13. Secondary Trend Selection
- 3.14. Home Screen Audio Alarm Settings
- 3.15. Settings and Maintenance
- 3.16. Info Options
- 3.17. Quick Reference Checklist
- 4. Operating the BIS™ Advance Monitoring System
- 4.1. Preparing for Operation
- 4.2. Cases
- 4.3. Alarms and Messages
- 4.4. Chart Data
- 4.5. Menu Map
- 4.6. Institutional Settings
- 5. Data Storage, Transfer, and Export
- 5.1. Cybersecurity and Data Integrity
- 5.2. Monitor Data Memory
- 5.3. BISx Data Memory
- 5.4. Types of Exported Data
- 5.5. Live Case Export
- 5.6. Saved Cases Export
- 5.7. Recording a Snapshot Event
- 5.8. BIS™ Saved Data Files
- 5.9. Logs Export
- 5.10. Viewing and Printing Saved Data in PDF Format
- 6. Service and Maintenance
- 6.1. Introduction
- 6.2. Servicing the BIS™ Advance Monitor
- 6.3. Battery Handling
- 6.4. Cleaning the BIS™ Advance Monitor
- 6.5. Instrument Identification
- 7. Diagnostics and Troubleshooting
- 7.1. Diagnostics
- 7.2. Troubleshooting
- 7.3. Administrator Mode
- 7.4. Demo Mode
- 7.5. Glossary
- 7.6. Specifications
- 7.6.1. General Specifications
- 7.6.2. Operating Environments
- 7.6.3. Power Supply
- 7.6.4. Battery Specifications
- 7.6.5. Controls and Connectors
- 7.6.6. Alarms
- 7.6.7. Display
- 7.6.8. Sound Pressure Data
- 7.6.9. BIS Specifications
- 7.6.10. EEG Specifications
- 7.6.11. BISx Module Specifications
- 7.6.12. Software/GUI Specifications
- 7.7. Electromagnetic Compatibility Specifications
- 7.8. Product Compliance
- 7.9. BIS™ Advance Components, Accessories, and Documentation
- 7.10. Warranty
- 7.11. Software License Agreement
Home Screen Numeric Section
112
BIS™ Advance Monitor
Label Function Description
1 ST numeric ST numeric in minutes and seconds (and
hours, if required)
2 ST unit markers If ST is over 59 minutes and 59 seconds,
an hour section will appear as well
3.10.7. SEF (Spectral Edge
Frequency)
SEF, the Spectral Edge Frequency (indicated in Hz) , is displayed as a numeric in the
numeric section and as a white line on the DSA graph. If no SEF range is set, the
alarm disabled icon will appear; if an SF range is set, the alarm range will appear and
an alarm will be triggered if the SEF is too high or too low.
For 4-channel mode, the Left SEF Number or Right SEF Number shall be displayed,
according to the user's selection of the brain hemisphere side.
Figure 42. SEF Display on Home Screen
2
1
4
3