Label Function Description
1 Burst Count numeric Burst Count numeric
2 Burst units Units in which Burst Count is displayed
3.10.5. SR (Suppression Ratio)
SR is displayed as a numeric in percent units, and, if selected as a secondary trend on
the BIS trend graph, also as a solid purple line on the BIS trend graph (0-100%).
If an SR alarm limit is set, the SR limit will be displayed as a dashed purple line on the
BIS trend graph, and the SR limit will be displayed near the SR Label. The SR limit is an
SR high limit; if SR surpasses the level set in the alarm limit, an alarm will be triggered.
For 4-channel monitoring, two SR Numbers, for the left and right hemispheres, will
be displayed.
Figure 40. SR Display on Home Screen
1 2
3
4