s User guide
INDICATIONS FOR USE
Surgical and dental instruments are designed to perform specific functions such as cuing, grasping, clamping, dissecting,
probing, retracting, draining, aspirating, suturing, or ligating. Surgical instruments may also be used to facilitate the insertion
of surgical implants.
CONTRAINDICATION
Instruments should not be used for anything other than their intended use.
WARNING
If any device is/was used in a patient with, or suspected of having Creutzfeld Jakob Disease (CJD), the device cannot be reused
and must be destroyed aer exposure as the instruments have not been validated to withstand the chemical and thermal
exposure recommended to eradicate prions.
Consult individual national infection control/prevention protocols for specific guidance regarding processing medical devices
with suspected exposure to CJD.
PRECAUTIONS
• IntegraMiltexinstrumentsaresuppliednon-sterile,unlessotherwisenotedandmustbecleaned,lubricatedand
sterilized prior to use according to hospital protocol and procedures outlined in this document.
• Inappropriateuseofinstrumentsmayresultinpatientinjury,damagedorbrokeninstruments.
• Propercleaning,handling,sterilizationandstandardroutinemaintenance(suchassharpening,ifapplicable)
will ensure that the instruments perform as intended and will extend their useful life.
• Delicatesurgical/dentalinstrumentsrequirespecialhandlingtopreventdamagingthetips.Usecaution
during cleaning and sterilization.
• Donotusedryheatsterilizationorexposetophenolsoriodophors.
• Donotapplyexcessivestressorstrainatjoints;misusewillresultinmisalignmentorcracksattheboxlocks
orjaws.
• Rongeursandbonecuingforcepsshouldonlybeusedtocutbone,neverwireorpin.Donottwistorapply
excessive stress during use.
• Wearappropriateprotectivegloves,eyewearandclothingwhenhandlingbiologicallycontaminated
instruments.
• Instrumentsmanufacturedfromdierentmetalsorwithspecialcoating,shouldbeprocessedseparatelyto
avoidelectrolyticactionbetweenthedierentmetals.
• Beforeuse,inspecttheinstrumentsforpossibledamage,wearornon-functioningparts.Carefullyinspect
thecritical,inaccessibleareas,jointsandallmovableparts.
• Damagedordefectiveinstrumentsshouldnotbeusedorprocessed.
•
Sutureneedlesaredesignedforsingle-useandshouldnotbereprocessed.Followsharpssafetyguidelinesforpointofusedisposable.
DECONTAMINATION AND STERILIZATION PROCEDURES
PersonnelshouldfollowacceptedguidelinesasrecommendedinANSI/AAMIST79-Comprehensiveguidetosteamsterilization
andsterilityassuranceinhealthcarefacilitiesandISO17664.ForEthyleneOxidesterilization,followtheANSI/AAMIST41Ethyl-
eneOxidesterilizationinhealthcarefacilities:safetyandeectiveness,4thedition.Sterilizationofmedicaldevicesinformation
to be provided by the manufacturer for the processing of resterilizable medical devices. Instruction for reprocessing reusable
surgicalinstrumentsisprovidedaccordingtoISO17664.Thedecontaminationproceduredoesnotsterilizetheinstruments.
RefertoandprocesstheinstrumentsasoutlinedintheSTERILIZATIONsection.
CARE AND HANDLING
1. PRECLEANING:
• Keepinstrumentsmoistanddonotallowbloodand/orbodilyuidstodryontheinstruments.
• Removegrosscontaminantswithsteadystreamoflukewarm/coolwater(below110°F/43°C).Rinseeach
instrument thoroughly. Do not use saline or chlorinated solutions.
• Openjawsofhingedinstrumentsforcleaning.Givespecialaentiontojointsandserrations.Instruments
having more than one part or piece must be disassembled to expose all surfaces to the cleaning process.
Retain all parts to facilitate reassembly.
• Separatesharpsanddelicatesurgicalinstruments.Avoidprocessinginstrumentsofdierentmetallic
composition together.
• Keepebonizedinstrumentsseparatefromotherstainlesssteelinstrumentstoavoidscratchestoand
removal of the ebonized coating.
2. RINSE: Rinse instruments thoroughly under warm distilled or demineralized water.
3. CLEANING:
CLEANING PRECAUTIONS:
• Ifappropriate,disassemblesurgicalinstrumentspriortocleaningandsterilization.
• Donotsoakinstrumentsinhotwater,alcohol,disinfectantsorantisepticstoavoidcoagulationofmucus,
bloodorotherbodyuids.Donotexceedtwohourssoakinginanysolution.
• Donotusesteelwool,wirebrushes,pipecleanersorabrasivedetergentstoremovesoilasthesewill
damage the instrument and lead to corrosion.
• Microsurgical,platedanddelicateinstrumentsshouldbecleanedmanuallyandshouldnotbeprocessedin
an ultrasonic cleaner. Carefully protect the tips of delicate microsurgical instruments throughout the entire
cleaning and sterilization process.
• Topreservethesurfacecoatingofebonizedinstruments,keepebonizedinstrumentsseparatefromother
instruments and avoid mechanical cleaning and abrasive cleaners as these processes can scratch the
surface and remove the surface coating.
• Coloranodizedaluminuminstrumentsmaylosetheircolorthroughtheuseofconventional,mechanical
treatment processes.
A. MANUAL CLEANING:Topreventformationofbiolm,cleaningshouldoccurassoonaspossibleaer
instrumentation is used.
• Chooseacleaningsolutionappropriateforsurgical/dentalinstrumentsandfollowmanufacturer’s
instructions for use.
• TheuseofneutralpHdetergentsisrecommendedtoavoidcorrosion,piingandbreakage.
• Usingasmall,cleanhand-heldbrush,removesoilfromallsurfacesofinstrumentwhilefullyimmersedin
solution.
B. ULTRASONIC AND MECHANICAL CLEANING
• Forultrasoniccleaning,followmanufacturer’sspecicationsforwaterlevel,concentrationlevelsofcleaning
agent and temperature.
• Useanultrasoniccleanertoremovesoilfromhardtoreachsurfacessuchasgrooves,crevicesandmoving
parts aer gross soil has been removed.
• Openordisassembleinstrumentsasappropriate.
• Whenusingmechanicalwasher,makesureallinstrumentsstayproperlyinplaceanddonottouchoroverlap
each other.
• Donotallowebonizedinstrumentstocomeincontactwitheachotherorotherinstruments.
• Alwaysfollowthemanufacturer’sspecicationsforautomaticwasher-sterilizersanduseafree-rinsing,
low-sudsingdetergentwithaneutralpH(6.0-8.5).Duetovariationsinwaterquality,thetypeofdetergent
anditsconcentrationmayrequireadjustmentforoptimalcleaning.
4. RINSE AND DRY:Rinseallinstrumentsthoroughlywithdistilled,pyrogen-freewatertoremovealltracesofdebrisandcleans-
ingagents.Makesureallinternallumensandratchetsarethoroughlyrinsed.
Instrumentsmustbethoroughlydriedandallresidualmoisturemustberemovedbeforetheyarestored.Useaso,absorbent
towel/cloth to dry external surfaces. Compressed air may be used to aid the drying process.
5. LUBRICATE:Theuseofawatersolubleinstrumentlubricantthatiscompatiblewiththemethodof
sterilization to be used is recommended before instruments are sterilized.
• Aerthoroughlycleaninginstruments,properapplicationoflubricantstoalljointsandmovablemating
surfaces will keep them moving freely and aid in protecting surface from mineral deposits.
• Properlubricationisrequiredforallinstruments,regardlessofsurfacecoatings.
• Notethatultrasoniccleanersremovealllubrication;therefore,thismaintenanceprocedureshouldbedone
routinely aer ultrasonic cleaning and before sterilization.
• Iffollowingthelubricatingprocedure,donotrinseaerthisstep.
6. STERILIZATION:
Aer following decontamination recommendations, reusable instruments are ready for sterilization.
• SeeANSI/AAMIST79.
• AAMIstandardsrecommendthatsterilizermanufacturer’swrieninstructionsforcycleparametersshould
alsobefollowed.Medicaldevicemanufacturer’sexposuretimestosterilizationtemperaturemayneedtobe
longer than minimum indicated by sterilizer manufacturer but must never be shorter. It is the responsibility of
the user to establish whether sterilizer meets these minimum recommendations.
• Instrumentsmaybepackagedinrigidcontainers,orpackagingclearedforuseinsterilization.Packaging
should ensure sterility of instruments until opened for use at the sterile field and permit removal of contents
without contamination.
• RecommendedsteamsterilizationparametertoachieveSterilityAssuranceLevel(SAL)of10
-6
:
Independentlaboratorytesting,referencingFDA(21CFRPart58)GoodLaboratoryPracticeRegulationshasvalidatedthe
following parameters.
High Vacuum Steam (Pre-vacuum)
Unwrapped in containers
Perforated stacked/unstacked
Solid unstacked
270ºF(132ºC) 4minutes
minimum
20minutes
Gravity Steam
Unwrapped in containers
Perforated unstacked
250ºF(121ºC) 45minutes 20minutes
Gravity Steam
Wrapped in cassees 250ºF(121ºC) 30minutes 20minutes
Ethylene Oxide (EtO)
Unwrapped in containers
Perforated unstacked
Solid unstacked
131ºF(55ºC) 60minutes(minimum)
Concentration:725mg/l
Humidity:70%
Aerationtime:8hoursminimum
IndependentlaboratorytestingconductedaccordingtoISO17664,hasvalidatedthefollowing
sterilization parameters.
Sterilization Method Temperature Exposure Time Dry Time
High Vacuum Steam (Pre-vacuum)
Wrapped 270ºF(132ºC) 3minutesminimum N/A
RETURNED GOODS POLICY
Productsmustbereturnedinunopenedpackageswithmanufacturer’ssealsintacttobeacceptedfor
replacement or credit unless returned due to a complaint of product defect. Determination of a product defect will be made by Integra
Miltex.Productswillnotbeacceptedforreplacementiftheyhavebeeninthepossessionofthecustomerformorethan90days.
REPAIRS AND MAINTENANCE
Improper,ineectiveandinsucientmaintenancecanreducethelifeofaninstrumentandwillinvalidatetheinstrument’swarranty.
Shouldinstrumentsrequirerepairormaintenance,contactIntegraMiltexforreturnauthorizationandaddress.
InstrumentsreturnedtoIntegraMiltexforrepairmusthaveastatementtestifyingthateachinstrumenthasbeen
thoroughlycleanedandsterilized.Failuretosupplyevidenceofcleaninganddisinfectionwillresultina
cleaning charge and delayed processing of your instrument repair.
PRODUCTINFORMATIONDISCLOSURE
INTEGRAANDITSSUBSIDIARIES(“INTEGRA”)ANDMANUFACTUREREXCLUDEALLWARRANTIES,EXCEPTINTEGRA’S
APPLICABLESTANDARDWARRANTYWHETHEREXPRESSEDORIMPLIED,INCLUDINGBUTNOTLIMITEDTO,ANYIMPLIEDWARRANTIES
OFMERCHANTABILITYORFITNESSFORAPARTICULARPURPOSE.NEITHERINTEGRANORMANUFACTURERSHALLBELIABLEFORANY
INCIDENTALORCONSEQUENTIALLOSS,DAMAGE,OREXPENSE,DIRECTLYORINDIRECTLY
ARISINGFROMUSEOFTHISPRODUCT.NEITHERINTEGRANORMANUFACTURERASSUMENORAUTHORIZEANYPERSON
TOASSUMEFORTHEMANYOTHERORADDITIONALLIABILITYOR
RESPONSIBILITYINCONNECTIONWITHTHESEPRODUCTS.
SYMBOLS USED ON LABELING
Manufacturer
1
European Authorized Representative
Catalog Number
LotNumber
Caution:SeeWarningsandPrecautions
ConsultInstructionsforUse
ComplieswithMDD93/42/EEC
EC REP
LOT
Federal(USA)lawrestrictsthisdevicetosaleby
or on the order of a physician or practitioner
Integra
®
Miltex
®
Care and Handling of Surgical and
Dental Instruments
DirectionsforUse/English
Integra,theIntegralogo,andMiltexareregisteredtrademarksofIntegraLifeSciences
CorporationoritssubsidiariesintheUnitedStatesand/orothercountries.
©2012IntegraLifeSciencesCorporation.AllRightsReserved. STANDARDIFURev. E 06/13
1
Company responsible for a device marketed under its own name regardless of whether
“manufacturedfor”or“manufacturedby”thecompany.
ComplieswithMDD93/42/EEC
IntegraYorkPA,Inc.
589DaviesDr.
York,PA17402USA
866-854-8300
717-840-9335
www.integralife.com/integra-miltex
EC REP
IntegraLifeSciencesServices(France)SASImmeubleSequoia2
97alléeAlexandreBorodine
ParcTechnologiquedelaPortedesAlpes
69800SaintPriest–France
Telephone:33(0)437475910
Fax:33(0)437475929
2
Do not reuse