Users Manual Part 8

348
Appendix
Controller Specications
Size: 5.67" high x 2.66" wide x 0.49" deep (143.92 mm x 67.57 mm x 12.33 mm)
Weight: 5.82 oz (165 grams)
Screen active area: 2.21" wide x 4.75" high (56.16 mm x 120.58 mm)
Operating temperature range: 41°F to 104°F (5°C to 40°C)
Storage temperature range: 32°F to 86°F (0°C to 30°C)
Operating relative humidity range: 20% to 90%, non-condensing
Storage relative humidity range: 20% to 90%, non-condensing
Operating atmospheric pressure: 700 hPA to 1060 hPA
Storage atmospheric pressure: 700 hPA to 1060 hPA
Communication distance: e Controller and Pod should be:
At startup: Adjacent and touching, with the Pod either in or out of tray, to
ensure proper communication during priming.
During normal operation: Within 5 feet (1.5 m) of each other. Depending on
the location, the communication distance may handle separations up to 50
feet (15 meters) away.
Alarm type: Audible. Output: ≥ 45 db(A) at 1 meter
IP (Ingress Protection) rating for moisture and dust: IP22 (protected from
touch by ngers and objects 12.5 millimeters or larger; not well-protected from
water - avoid liquid)
Notication type: Audible and vibratory
Battery: Rechargeable Li-ion battery, 3.8V, 2800 mAh
Battery life: Full charge covers approximately 2 days of typical use aer 2 years.
Controller Service Life: 2 Year s
Shelf Life (Starter Kit): 18 months
Battery charger operating line voltage: 100 to 240 VAC, 50/60 Hz
Only use the Noetic approved power adapter (Insulet PN PT-000428) with
the Controller.
Dexcom Specications
For information about Dexcom operating specications, see the Dexcom G6
CGM System User Guide.