Users Manual Part 8

359
Appendix
Electromagnetic Immunity
Electrical Fast
Transient/burst
(IEC 61000-4-4)
± 2 kV power
supply lines
± 2 kV Input
DC Power
Port
± 1 kV
input/output
lines
± 2 kV
power
supply lines
± 2 kV
Input DC
Power Port
± 1 kV
input/
output lines
Mains power quality should be that of a
typical domestic, commercial, or hospital
environment.
Surge (IEC
61000-4-5)
± 1 kV
dierential
mode
± 2 kV
common
mode
± 1 kV
dierential
mode
± 2 kV
common
mode
Mains power quality should be that of a
typical domestic, commercial, or hospital
environment.
Conducted
Disturbances
induced by
RF elds (IEC
61000-4-6)
3V 150 KHz-
80 MHz
6V in ISM
and amateur
radio bands
between 150
KHz and 80
MHz
3V 150
KHz-80
MHz
6V in
ISM and
amateur
radio bands
between
150 KHz
and 80 MH
Suitable for most environments. Keep
portable RF communications equipment
at least 12 inches (30 cm) away from the
Omnipod 5 System.