Users Manual Part 8
358
Appendix
Electromagnetic Emissions
is equipment is intended for use in the electromagnetic environment
specied below. e user of this equipment should ensure that is used in such an
environment.
Emissions Compliance
according to
Electromagnetic environment
RF Emissions
(CISPR11)
Group 1 e Pod, Controller, and CGM emit low-
level electromagnetic energy (RF) in order
to communicate. Although unlikely, nearby
electronic equipment may be aected.
CISPR B
Emissions
Classication
Class B e System is suitable for use in all establishments,
including domestic establishments.
Harmonic
Emissions
(IEC 61000-3-2)
Class A
Voltage
Fluctuations/
Flicker Emissions
(IEC 61000-3-3)
P
st
≤ 1.0
P
lt
≤ 0.65
d
c
≤ 3%
d
max
≤ 4%
d
(t)
≥ 200 ms
during a voltage
change should be
≤ 3%
Electromagnetic Immunity
e System is intended for use in the electromagnetic environment specied
below. You should observe these requirements in the use of the System.
Immunity
against
IEC 60601-1-2
test level
Compliance
level (of this
device)
Electromagnetic environment
ElectroStatic
Discharge, ESD
(IEC 61000-4-2)
contact
discharge:
± 8 kV
air discharge:
± 15 kV
± 8 kV
± 15 kV
If oors are covered with synthetic material,