Users Manual Part 7
Automated Mode Alarms 24
306
24 Automated Mode Alarms
24.1. Advisory Alarm List
Advisory alarms inform you of a situation that needs your attention in the near
future.
Automated Delivery Restriction
Only occurs in Automated Mode.
Omnipod 5 App Screens:
Lock Screen:
Cause
Insulin was either paused for too long or at maximum
delivery for too long while the Omnipod 5 System was in
Automated Mode.
Tone (Pod)
• 6 beep tone, repeats once every minute for 3 minutes
• Pattern repeats every 15 minutes
Vibration/Tone
(Controller or
smartphone)
• 3 second tone
• 3 second vibration
• Vibration and tone repeat every 15 minutes until
acknowledged