Users Manual Part 7

Activity Feature 23
302
23 Activity Feature
23.1. About the Activity Feature
Warning: ALWAYS monitor for symptoms of hypoglycemia while the
Activity feature is enabled. Hypoglycemia can still occur when using the
Activity feature. Follow your healthcare provider’s advice on hypoglycemia
avoidance and treatment. If untreated, hypoglycemia can lead to seizure, loss
of consciousness or death.
While in Automated Mode, you cannot start a temp basal or manually pause
insulin delivery. e Omnipod 5 System provides an option for modied
automated insulin delivery through the Activity feature. e Activity feature
can be useful in times when you need less insulin, for example, when you are
exercising.
While Activity is enabled, the Omnipod 5
System does the following:
Reduces automated insulin delivery
Sets your Target Glucose to 150 mg/dL,
regardless of your target settings
With Activity enabled, you can still deliver a
bolus as you normally would.
Activity can be set for a duration of 1-24
hours, in increments of 1 hour. You
can cancel Activity at any time. Upon
cancellation or expiration of the dened
time period, full automated insulin
delivery starts on its own and SmartAdjust
technology returns to using the Target
Glucose dened in your settings.
e Activity feature ends if the Pod
is deactivated. You need to re-enter
Automated Mode and then enable Activity
with your new Pod.
Talk to your healthcare provider about the timing of starting the Activity feature to
address your anticipated period of decreased insulin needs.
Note: In the event of a loss of Pod and CGM communication and the
Omnipod 5 System enters Limited state, Activity remains enabled.
Note: You may see an increase in your displayed IOB when the Activity feature
starts and a decrease in your IOB when the Activity feature time period ends
because of the way insulin is calculated.