Users Manual Part 5
201
Alarms, Action and Reminder Notications 1313 Alarms, Action and Reminder Notications
Missed Bolus
Screen Displayed
Cause
Reminds you that you have not delivered a bolus within
the time period you have specied.
Pod sound:
None
Controller/
Smartphone
sound and
vibration:
Every 15 minutes until acknowledged.
What to do
Acknowledge the message (see "Acknowledging
reminder notications" on page 198). Consider your
meal schedule.
Note: To edit these reminders, see "Missed Bolus" on page 236.
Check BG Aer Pod Change
Screen Displayed
Cause
Reminds you to check your blood glucose and the
cannula infusion site 90 minutes aer activating a new
Pod.
Pod sound:
None
Controller/
Smartphone
sound and
vibration:
Repeats once, 15 minutes aer initial notication
unless acknowledged.
What to do
Acknowledge the message (see "Acknowledging
reminder notications" on page 198). Check your
blood glucose. Check the infusion site to see if the
cannula is properly inserted.
Note: If there is a valid CGM value present at the time, in either Manual or
Automated Mode, this alert will automatically be cleared by the Omnipod 5 App.