Users Manual Part 4

171
Alarms, Action and Reminder Notications 1313 Alarms, Action and Reminder Notications
13.7. Advisory Alarm List
Caution: ALWAYS respond to Pod Expired, Low Pod Insulin, and Pod
Shut-O Advisory Alarms when they occur. ese alarms escalate to Hazard
Alarms if no action is taken. When Hazard Alarms occur, insulin delivery
stops.
Advisory alarms inform you of a situation that needs your attention in the near
future.
Low Pod Insulin
Screen Alert Description
Omnipod 5 App: Why it occurs: e amount of insulin in
the Pod is below the value specied in your
settings.
Pod sound:
8 beep tone pattern
Tone pattern sounds once every 3
minutes for 60 minutes
Controller/Smartphone sound and
vibration:
3 second tone
3 second vibration
Vibration and tone repeat every 15
minutes until acknowledged
What to do:
1. Tap OK.
2. Change your Pod.
Lock Screen:
Note: is escalates to the Pod Out of Insulin hazard alarm if ignored.
Note: To change this value, see "Low Pod Insulin" on page 128.