Users Manual Part 4

161
Alarms, Action and Reminder Notications 1313 Alarms, Action and Reminder Notications
important safety alarms. You cannot use the App until Do Not Disturb permission
is provided and Notications is turned ON.
To respond to a hazard alarm or advisory alarm:
1. Wake up your Controller or smartphone. e Lock screen shows an alarm
message along with the hazard alarm ( ) icon or advisory alarm ( ) icon.
2. Aer unlocking your Omnipod 5 App, follow the on-screen instructions or
see the individual alarm details starting on page 162.
Note: You can use your Omnipod 5 System even if you do not address an
advisory alarm immediately. However, you must acknowledge a hazard alarm
before you can use your Omnipod 5 System.
Tip: If you follow the Omnipod 5 App's instructions and are still not able to
silence a hazard alarm, see "13.9. Silencing Unresolved Alarms" on page 197.
Note: If a temp basal or extended bolus is running when a Pod hazard alarm
occurs, the Omnipod 5 App informs you that it was canceled.
Timing of Pod Alarms on the Omnipod 5 App
If the Pod is sounding a hazard alarm, the Pod sends a signal to your
Omnipod 5 App.
If the Controller or smartphone running the Omnipod 5 App is in range and
awake within 25 seconds of the Pod's initial alarm sound, your Omnipod 5
App also sounds an alarm and displays the alarm message.
If the Controller or smartphone running the Omnipod 5 App is in range
but asleep, there could be a delay of up to 6 minutes between when the Pod
sounds the alarm and when your Omnipod 5 App sounds the alarm.
If the Controller or smartphone running the Omnipod 5 App is out of
range of the Pod, your Omnipod 5 App cannot receive any communication
from the Pod. erefore, if you hear a Pod alarm or notication, bring your
Controller or smartphone in range of the Pod and wake up your Omnipod
5 App. Within 25 seconds, your Omnipod 5 App sounds the alarm and
displays the alarm message.
Diagnostics
Warning: DO NOT continue using an activated Pod that fails to beep
during a diagnostic test. e Pod should be changed immediately. If the
Omnipod 5 App fails to beep during a diagnostic test, call Customer Care
immediately. Continuing to use the Omnipod 5 System in these situations
could put your health and safety at risk.