Users Manual Part 3

Browsing Your History and Records 11
140
11 Browsing Your History and Records
11.5. Calculations for History Summaries
is section lists the calculations for the summary data shown on the History
screens.
Glucose Summaries
e calculations used for the glucose summaries include CGM values (including
HIGH and LOW values) and exclude manually-entered readings.
Item Calculation
Row 1
Average
CGM
=
Sum of all CGM values
Total number of CGM values
Note: HIGH CGM values are included as 401 and LOW CGM
values as 39.
CGM in
Range
=
Qty of CGM values within Glucose Goal Range
x 100
Total number of CGM values
CGM
Above
=
Qty of CGM values above Glucose Goal Range upper limit
x 100
Total number of CGM values
CGM
Below
=
Qty of CGM values below Glucose Goal Range lower limit
x 100
Total number of CGM values
Insulin Delivery Summaries
Bolus insulin calculations include:
SmartBolus Calculator boluses
manual boluses
any partial bolus amounts delivered from immediate or extended boluses
that were canceled that the Pod can conrm
Basal insulin calculations include:
basal adjustments from SmartAdjust technology (Automated Mode)
Basal Programs (Manual Mode)
any adjustments for temp basals, pause periods, or times without an active
Pod