Users Manual Part 3

105
Temporary Basal Rates and Presets 77 Temporary Basal Rates and Presets
7.2. Starting a Temp Basal
Note: You cannot start or cancel a temp basal during an immediate bolus, but you
can start or cancel a temp basal while an extended bolus is in progress.
To start a temp basal:
1. Navigate to:
Menu button ( ) > Set Temp Basal
e screen shows a graph of the Basal Program in progress.
2. Tap the Basal Rate eld and scroll to the desired change in the basal rate:
If using a percent (%) change:
An UP ARROW ( ) indicates increasing the basal rate above that of
the Basal Program in progress.
A DOWN ARROW ( ) indicates decreasing the basal rate below that
of the Basal Program in progress.
If using a at rate (U/hr), scroll to select the basal rate for the entire temp
basal period.
Note: To change whether temp basals are congured as percent (%) or U/hr,
see "10.3. Basal and Temp Basal Settings" on page 132.
Note: e scroll wheel will not scroll above your Maximum Basal Rate. To
adjust your Maximum Basal Rate, see "Maximum Basal Rate" on page 132.
Tip: You can turn OFF insulin delivery for the duration of the temp basal
by setting a decrease of 100% or setting the temp basal to 0 U/hr. For more
information, see "Temp basal limitations" on page 111 and "7.7. Temporary
Basal Rates Delivery" on page 109.
3. Tap the Duration eld and scroll to the desired temp basal duration
(between 30 minutes and 12 hours).