Users Manual Part 2
85
Activating and Changing Your Pod 55 Activating and Changing Your Pod
Fill the Pod with Insulin
To ll the Pod with insulin (screen step )
1. Locate the arrow on the underside of the
Pod. e arrow points to the insulin ll port.
Tip: You can leave the Pod in its tray during
lling and activating.
2. Insert the ll syringe straight down—not at
an angle—into the ll port.
3. Depress the ll syringe plunger to transfer
the insulin into the Pod.
Listen for two beeps from the Pod during the
lling process (screen step ):
4. Be sure to completely empty the ll syringe,
even aer hearing the two beeps.
Note: e Pod must contain a minimum
of 85 units of insulin to function. e Pod
beeps twice aer it has been lled with 85
units of insulin. If you lled the Pod with more than 85 units and still do not
hear the two beeps, contact Customer Care.
Note: Aer lling the Pod, continue to the next step immediately. If two
hours pass before activating the lled Pod, the Pod becomes unusable.
5. Remove the needle from the insulin ll port. e port is self-sealing; insulin
will not leak aer the needle is removed.
6. Discard the ll needle in a sharps container.
Activate the Pod
To activate the Pod:
1. Place the Controller or smartphone
running the Omnipod 5 App next to the
Pod so they are touching. e Pod should
be in its plastic tray during this process.
Note: ALWAYS ensure that no other
Pods are being activated within 20 feet
(6 m) of your Omnipod 5 App before
lling a Pod. If the App detects more
than one Pod, you will not be able to
proceed.
2. Tap NEXT.
5.4. Filling, Activating, Applying, and Starting the Pod
Warning: NEVER use a Pod if, while you are lling the Pod, you feel
signicant resistance while pressing the plunger down on the ll syringe. Do
not try to force the insulin into the Pod. Signicant resistance may indicate
that the Pod has a mechanical defect. Using this Pod could result in under-
delivery of insulin that can lead to hyperglycemia.
Warning: NEVER inject large bubbles or pockets of air when lling the Pod
with insulin. Air in the system takes up space where insulin should be and
can aect insulin delivery. Doing so could result in over-delivery or under-
delivery of insulin, which can lead to hypoglycemia or hyperglycemia.
Caution: ALWAYS insert the ll syringe into the ll port and not into any
other location on the Pod. Do not insert the ll syringe more than once into
the ll port. Use only the ll syringe and needle that came with your Pod. e
ll syringe is intended for single use only and should only be used with the
Omnipod 5 System. Failure to follow the instructions above may result in
damage to your Pod.