5 Activating and Changing Your Pod
5.4. Filling, Activating, Applying, and Starting the Pod
Warning: Before lling a Pod, ensure that no other Pods are being activated
within 20 feet (6 m) of your Omnipod 5 app.
Warning: NEVER use a Pod if you feel resistance when you depress the
plunger. is condition can result in interrupted insulin delivery.
Warning: NEVER inject air into the ll port. Doing so may result in
unintended or interrupted insulin delivery.
Caution: Be sure to insert the ll syringe into the ll port and not any other
location on the Pod. Do not insert the ll syringe more than once into the ll port.
Only use the ll syringe and needle that came with your Pod. e ll syringe is
intended for single use only and should only be used with the Omnipod 5 System.
Fill the Pod with insulin
To ll the Pod with insulin (screen step )
1. Locate the arrow on the underside
of the Pod. e arrow points to the
insulin ll port.
Tip: You can leave the Pod in its tray
during lling and activating.
2. Insert the ll syringe straight down—
not at an angle—into the ll port.
3. Depress the ll syringe plunger to
transfer the insulin into the Pod.
Listen for two beeps from the Pod during
the lling process (screen step
):
4. Be sure to completely empty the ll
syringe, even aer hearing the two
beeps.
Note: e Pod must contain a
minimum of 85 units of insulin in order to function. e Pod beeps twice
aer it has been lled with 85 units of insulin. If you lled the Pod with more
than 85 units and still do not hear the two beeps, please call Customer Care.
Note: Aer lling the Pod, continue to the next step immediately. If two
hours pass before activating the lled Pod, the Pod becomes unusable.
5. Remove the needle from the insulin ll port. e port is self-sealing; insulin
will not leak aer the needle is removed.
6. Discard the ll needle in a sharps container.