Activating and Changing Your Pod 5
Before activating a Pod,
1. Gather the necessary supplies:
A vial of rapid-acting U-100 insulin cleared for use in the Omnipod 5
System. See "1.4. General Warnings and Safety Information" on page
8 for a list of the approved insulin types that can be used with the
Omnipod 5 System.
An unopened Omnipod 5 Pod
Alcohol prep swabs
Controller or smartphone with Omnipod 5 app
2. Wash your hands before starting and keep them clean throughout the Pod
change process.
3. Check the insulin for signs of deterioration according to the manufacturer's
instructions for use.
4. Check the Pod's packaging for damage. If undamaged, open it and inspect the
Pod for signs of damage.
5. If the insulin or Pod is below 50°F (10°C), allow it to warm up to room
temperature before proceeding.
Conrm that you are using an Omnipod 5 Pod prior to beginning Pod activation.
Look for the Omnipod 5 logo on the Pod tray lid and the clear needle tab.