Use Instructions

300
Appendix
Electromagnetic Emissions
CISPR B
Emissions
Classication
Class B
e System is suitable for use in all establishments
including domestic establishments.
Electromagnetic Immunity
e System is intended for use in the electromagnetic environment specied
below. You should observe these requirements in the use of the System.
Immunity
against
IEC 60601-1-2
test level
Compliance
level (of this
device)
Electromagnetic environment
ElectroStatic
Discharge, ESD
(IEC 61000-4-2)
contact
discharge:
± 8 kV
air discharge:
± 15 kV
± 8 kV
± 15 kV
If oors are covered with synthetic material,
try to avoid electrostatic discharges.
Power
frequency
magnetic elds
50/60 Hz (IEC
61000-4-8)
30 A/m 400 A/m
Suitable for most environments. Magnetic
eld strengths in excess of 400 A/m would
be unlikely except in close proximity to
industrial magnetic devices.
Radiated RF
(IEC 61000-4-3)
10 V/m at
80 MHz–
2.7 GHz
10 V/m
Suitable for most environments. Keep
portable RF communications equipment
at least 12 inches (30 cm) away from the
Omnipod 5 System.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflections from structures, objects, and people.
a
Field strengths from xed transmitters, such as base stations for radio (cellular/
cordless) telephones and land mobile radios, amateur radio, AM and FM radio
broadcast, and TV broadcast, cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to xed RF transmitters, an electromagnetic site
survey should be considered. If the measured eld strength in the location in which the
equipment is used exceeds the applicable RF compliance level above, the equipment
should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the equipment.
Effective Date: 22DEC2020, PCO-000267