Use Instructions
287
Appendix
Summary of Settings and Options
e options for the various Omnipod 5 Automated Insulin Delivery System
settings are:
Time format 12-hour
Time zone GMT-11:00 to GMT+13.00.
Daylight Savings Time ON or OFF. Default based on date and time zone.
Date format MM/DD/YYYY
Screen time-out 30, 60, 120 seconds. Default is 30 seconds.
PIN 4 digits from 0 to 9.
CGM G6 transmitter
serial number
6 characters.
Maximum Basal Rate 0.05-30 U/hr. Default is 3.00 U/hr.
Basal rate Units/hr. Range: 0 U/hr to Maximum Basal Rate in
0.05 U/hr increments.
Basal Programs Maximum of 12.
Basal rate segments 24 per Basal Program.
HypoProtect™ Range: 1 - 72 hrs
Temp basal %, units/hr, or OFF. Default is OFF.
Duration: 30 min to 12 hrs in 30-min increments.
Temp basal (set to %) Range: 100% decrease (0 U/hr) to 95% increase from
current basal rate in 5% increments. Cannot exceed
Maximum Basal Rate.
Temp basal (set to U/hr) Range: 0 U/hr to Maximum Basal Rate in increments of
0.05 U/hr.
Temp basal presets Maximum of 12.
BG/CGM Goal Range
(for blood glucose
history)
Lower and upper limits: 70 to 200 mg/dL in 1 mg/dL
increments .
BG reminder ON or OFF. Default is OFF.
Maximum of 4 active at one time.
Reminder can occur between 30 min and 4 hrs aer
bolus is started. Set in 30-minute increments.
Custom reminder Maximum of 4. Set to Daily, One time only, OFF.
Target BG value Maximum of 8 segments; 110 to 150 mg/dL in
10 mg/dL increments.
Effective Date: 22DEC2020, PCO-000267