Product Info
203
Alarms, Action and Reminder Notications20
Missing CGM Va lu e s
Only occurs in Automated Mode.
Cause
e Pod has not received
CGM values for more
than one hour. e system
will continue to operate in
Automated Mode: Limited
state until CGM values are
received or until you switch
to Manual Mode.
Vibration/Tone
e Omnipod 5 app sounds
the alarm.
What to do
¾ Tap OK to acknowledge
the alert.
For more information about Automated Mode: Limited state, see "15.5. About
Automated Mode: Limited" on page 144.
Check your
Dexcom G6 app to see if there are CGM values present or if
the cause of the loss of communication is related to the sensor/transmitter.
Examples to look for within the
Dexcom G6 app include sensor error/
expiration, transmitter error/expiration, sensor warm-up, or signal loss alert.
If the
Dexcom G6 app is receiving readings, there may be a temporary
communication issue between your Pod and the
Dexcom G6. You may decide
to switch to Manual Mode or wait for a CGM value to be received while in
Automated Mode: Limited state. If this is occurring frequently, check to see if
the Pod and CGM are located on your body within line of sight. If not, when
you remove one, be sure to position the new one so that your Pod and CGM are
within line of sight.
Note: For information about your DexcomG6 app, refer to your DexcomG6
User Guide.
Pod Expired
Cause
Your Pod will stop delivering
insulin soon.
Vibration/Tone
e Pod sounds the alarm.
e Omnipod 5 app also
sounds the alarm, if enabled.
What to do
1. Tap OK.
2. Change your Pod.
Note: is escalates to the Pod Expired hazard alarm if ignored.
Trouble-
shooting
Effective Date: 22DEC2020, PCO-000267