Product Info

201
Alarms, Action and Reminder Notications20
While in Manual Mode, you can check your CGM graph to nd out whether your
insulin has been paused or has been at a maximum for a long time.
Automated Mode: Limited state is the automated insulin delivery used when
CGM values are not available. Insulin delivery is based on your settings and recent
history. At this point, you are receiving a steady basal rate that will not vary until
CGM readings are restored. Basal insulin could remain paused if automated
insulin was paused before entering Automated Mode: Limited state.
Aer 5 minutes of Manual Mode, you can return to Automated Mode once you
have conrmed your CGM readings are accurate.
For more information about Automated Mode: Limited state, see "15.5. About
Automated Mode: Limited" on page 144.
Low battery
Cause
e battery charge has 15%
or less remaining.
Vibration/Tone
None. To preserve the
battery, the Omnipod 5 app
is silent.
What to do
1. Tap OK.
2. Recharge your controller
as soon as possible.
Warning: If the battery state of charge becomes critically low, the controller
will turn itself OFF. At this point, you cannot use the controller until you have
plugged in the charger to recharge the battery.
Caution: ere is no controller low battery hazard alarm when the battery
completely runs out. Plug in the charger as soon as possible aer seeing the low
battery message.
Note: If your controller or smartphone has turned OFF due to a critically low
battery, your Pod continues to deliver Automated Mode insulin or Manual
Mode basal insulin according to the Basal Program in progress or temp basal.
If you do not charge your controller or smartphone, this insulin delivery
continues until the Pod expires. When the controller or smartphone has turned
OFF due to a critically low battery, your Pod will continue to sound alarms and
notications.
Note: You can use your controller or smartphone while it is charging.
Tip: Develop a routine to charge the controller or smartphone at the same time
every day. Do not wait for the low battery message.
Trouble-
shooting
Effective Date: 22DEC2020, PCO-000267