Product Info

200
20Alarms, Action and Reminder Notications
20.8. Advisory Alarm List
Advisory alarms inform you of a situation that needs your attention in the near
future.
Warning: ree advisory alarms (Pod Expiration, Low Pod Insulin, and Pod
Shut-O) become hazard alarms and result in a stoppage of insulin delivery if
ignored. Be sure to respond to all advisory alarms when they occur.
Automated Delivery Restriction
Only occurs in Automated Mode.
Cause
Insulin was either
paused for too long or
at maximum delivery
for too long while the
Omnipod 5 app was in
Automated Mode.
What to do
1. Tap NEXT to see the next screen.
2. Use a BG meter to conrm your blood glucose.
If low conrmed, consider treatment.
If high conrmed, check infusion (Pod) site
and ketones.
If your CGM value is not what you expected,
you may need to calibrate or replace your
Dexcom G6 sensor.
3. Tap NEXT aer you conrm your blood glucose.
4. Tap SWITCH TO MANUAL MODE, then stay in
Manual Mode for ve minutes.
Vibration/Tone
e Pod sounds the
alarm. e
Omnipod 5
app also sounds the
alarm, if enabled.
Trouble-
shooting
Effective Date: 22DEC2020, PCO-000267