Product Info

156
17 Changing Settings
To verify that your Omnipod 5 app and Pod's alarms and vibration functions are
working properly, you can test them as follows:
1. Navigate to: Menu icon ( ) > Settings > PDM Device > Check Alarms.
2. If you have an active Pod, tap PAUSE INSULIN and tap YES.
3. Tap CHECK ALARMS to initiate the alarm check.
4. Listen and feel: e Omnipod 5 app beeps and vibrates three times. en, if
you are wearing a Pod, the Pod beeps several times and sounds the alarm tone
for several seconds.
5. If the Pod alarms did not work properly, tap NO. en either tap CHECK
AGAIN to retry testing the alarms, or tap DONE and change your Pod.
6. If the Omnipod 5 app alarms did not work properly, tap NO. en either tap
CHECK AGAIN to retry testing the alarms or call Customer Care.
7. If the beeps and vibrations worked properly, tap YES. If you paused insulin to
check the alarms, tap YES to start insulin delivery.
Warning: If the Omnipod 5 app fails to beep, call Customer Care
immediately. If an activated Pod fails to beep, change the Pod immediately.
Continuing to use the Omnipod 5 System in these situations may put your
health and safety at risk.
17.2. Pod Sites Settings
e Pod Sites setting controls whether a diagram of the body is available when you
activate a new Pod. You can use this body diagram during Pod activation to review
where you placed your recent Pods and to mark where you place the new Pod.
To turn the Pod Sites setting ON or OFF:
1. Navigate to: Menu icon ( ) > Settings > Pod Sites.
2. Tap the toggle to turn the Pod Sites setting ON or OFF. If you want the ability
to record your Pod site on the diagram, turn Pod Sites ON. If you do not want
to use this diagram, turn it OFF.
3. Tap SAVE.
Settings
History
Updates
Effective Date: 22DEC2020, PCO-000267