User Manual
Table Of Contents
- Your New OmniPod Insulin Management System
- Getting Started
- The OmniPod Starter Kit
- Set Up the Personal Diabetes Manager (PDM)
- The Setup Wizard
- Enter your PDM ID
- Select the ID screen color
- Set date and time
- Enter basal settings
- Enter blood glucose sound setting and BG goal
- Set the suggested bolus calculator
- Enter target blood glucose value
- Enter minimum BG allowed for use in bolus calculation
- Enter insulin-to-carbohydrate ratio
- Enter correction factor
- Set reverse correction
- Enter the duration of insulin action
- Select bolus increment and enter maximum bolus
- Set extended bolus doses
- Set low reservoir advisory
- Set expiration notification
- Understanding and Adjusting Basal Rates
- Understanding and Delivering Bolus Doses
- Using the Personal Diabetes Manager
- Checking Your Blood Glucose
- The Built-in FreeStyle® Blood Glucose Meter
- The FreeStyle® Blood Glucose Test Strips
- The FreeStyle® Control Solution
- Performing a Control Solution Test
- Performing a Blood Glucose Reading
- Blood Glucose Results and the Suggested Bolus Calculator
- Entering Blood Glucose Readings Manually
- Editing Tags
- Low and High Blood Glucose Readings
- Important Health-Related Information
- Understanding Your Records
- Living with Diabetes
- Alerts and Alarms
- Communication Failures
- Appendix
- Pod Care and Maintenance
- Personal Diabetes Manager Care and Maintenance
- Storage and Supplies
- Suggested Bolus Calculator Examples and Guidelines
- OmniPod System Options and Settings
- Pod Specifications
- Accuracy Test Results
- Personal Diabetes Manager Specifications
- Blood Glucose Meter Specifications
- OmniPod System Label Symbols
- Personal Diabetes Manager Icons
- OmniPod System Notice Concerning Interference
- Electromagnetic Compatibility
- Customer Bill of Rights
- Limited Warranty for the Personal Diabetes Manager
- HIPAA Privacy Notice
- Glossary
- Index
Understanding and Delivering Bolus Doses
33
4
4. If you have set the blood glucose reminder option to On a
nd
you want to add a reminder:
a. Press Yes.
b. Enter the time for the reminder, then press OK.
5. Press Confirm to begin the bolus delivery.
Deliver a bolus manually (suggested bolus calc is Off)
1. On the Home screen, use the Up/Down Controller buttons
to choose Bolus, then press Select.
2. Enter the number of units you want to deliver, then press
Enter.
3. If the blood glucose reminders feature is turned On, you can
now set a blood glucose reminder.
4. Press Confirm to begin the bolus delivery.
If the reverse correction feature is turned On and if
your blood glucose level is below your target, the
System subtracts a correction amount from the meal
portion of the bolus.
If the reverse correction setting is turned Off, the
System does not subtract anything for a blood glu-
cose level below your target.
Figure 4-6
As a safety feature, the OmniPod System only allows
you to give a bolus at or below the maximum bolus
dose you have set. See Chapter 6, Using the Personal
Diabetes Manager, for information on resetting your
maximum bolus dose. If suggested bolus calc is On,
you can enter a temporary manual override to make
an adjustment above your max bolus limit (Figure 4-1
and Figure 4-2). Consult your healthcare provider