User Manual
Table Of Contents
- Your New OmniPod Insulin Management System
- Getting Started
- The OmniPod Starter Kit
- Set Up the Personal Diabetes Manager (PDM)
- The Setup Wizard
- Enter your PDM ID
- Select the ID screen color
- Set date and time
- Enter basal settings
- Enter blood glucose sound setting and BG goal
- Set the suggested bolus calculator
- Enter target blood glucose value
- Enter minimum BG allowed for use in bolus calculation
- Enter insulin-to-carbohydrate ratio
- Enter correction factor
- Set reverse correction
- Enter the duration of insulin action
- Select bolus increment and enter maximum bolus
- Set extended bolus doses
- Set low reservoir advisory
- Set expiration notification
- Understanding and Adjusting Basal Rates
- Understanding and Delivering Bolus Doses
- Using the Personal Diabetes Manager
- Checking Your Blood Glucose
- The Built-in FreeStyle® Blood Glucose Meter
- The FreeStyle® Blood Glucose Test Strips
- The FreeStyle® Control Solution
- Performing a Control Solution Test
- Performing a Blood Glucose Reading
- Blood Glucose Results and the Suggested Bolus Calculator
- Entering Blood Glucose Readings Manually
- Editing Tags
- Low and High Blood Glucose Readings
- Important Health-Related Information
- Understanding Your Records
- Living with Diabetes
- Alerts and Alarms
- Communication Failures
- Appendix
- Pod Care and Maintenance
- Personal Diabetes Manager Care and Maintenance
- Storage and Supplies
- Suggested Bolus Calculator Examples and Guidelines
- OmniPod System Options and Settings
- Pod Specifications
- Accuracy Test Results
- Personal Diabetes Manager Specifications
- Blood Glucose Meter Specifications
- OmniPod System Label Symbols
- Personal Diabetes Manager Icons
- OmniPod System Notice Concerning Interference
- Electromagnetic Compatibility
- Customer Bill of Rights
- Limited Warranty for the Personal Diabetes Manager
- HIPAA Privacy Notice
- Glossary
- Index
Appendix
153
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment
and the System
You can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communi-
cations equipment (transmitters) and the System, as recommended below, according to the maximum output power of the com-
munications equipment.
Rated maximum output power
of transmitter, in watts
Separation distance according to frequency of transmitter, in meters
150 kHz to 80 MHz, d = 1.17
√
P 80 MHz to 800 MHz, d = 0.35
√
P 800 MHz to 2.5 GHz, d = 0.7
√
P
0.01 0.12 0.035 0.070
0.1 0.37 0.11 0.22
1 1.17 0.35 0.7
10 3.70 1.11 2.21
100 11.7 3.5 7.0
For transmitters rated at a maximum output power not listed above, the recommended separation distances in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from struc-
tures, objects, and people.