User Manual
Table Of Contents
- Your New OmniPod Insulin Management System
- Getting Started
- The OmniPod Starter Kit
- Set Up the Personal Diabetes Manager (PDM)
- The Setup Wizard
- Enter your PDM ID
- Select the ID screen color
- Set date and time
- Enter basal settings
- Enter blood glucose sound setting and BG goal
- Set the suggested bolus calculator
- Enter target blood glucose value
- Enter minimum BG allowed for use in bolus calculation
- Enter insulin-to-carbohydrate ratio
- Enter correction factor
- Set reverse correction
- Enter the duration of insulin action
- Select bolus increment and enter maximum bolus
- Set extended bolus doses
- Set low reservoir advisory
- Set expiration notification
- Understanding and Adjusting Basal Rates
- Understanding and Delivering Bolus Doses
- Using the Personal Diabetes Manager
- Checking Your Blood Glucose
- The Built-in FreeStyle® Blood Glucose Meter
- The FreeStyle® Blood Glucose Test Strips
- The FreeStyle® Control Solution
- Performing a Control Solution Test
- Performing a Blood Glucose Reading
- Blood Glucose Results and the Suggested Bolus Calculator
- Entering Blood Glucose Readings Manually
- Editing Tags
- Low and High Blood Glucose Readings
- Important Health-Related Information
- Understanding Your Records
- Living with Diabetes
- Alerts and Alarms
- Communication Failures
- Appendix
- Pod Care and Maintenance
- Personal Diabetes Manager Care and Maintenance
- Storage and Supplies
- Suggested Bolus Calculator Examples and Guidelines
- OmniPod System Options and Settings
- Pod Specifications
- Accuracy Test Results
- Personal Diabetes Manager Specifications
- Blood Glucose Meter Specifications
- OmniPod System Label Symbols
- Personal Diabetes Manager Icons
- OmniPod System Notice Concerning Interference
- Electromagnetic Compatibility
- Customer Bill of Rights
- Limited Warranty for the Personal Diabetes Manager
- HIPAA Privacy Notice
- Glossary
- Index
Appendix
151
Electromagnetic Emissions
This equipment is intended for use in the electromagnetic environment specified below. The user of this equipment should assure
that is used in such an environment.
Emissions Compliance according to Electromagnetic environment
RF Emissions (CISPR 11) Group 2 The Pod and the PDM emit low level electromagnetic energy (RF) in
order to communicate. Although unlikely, nearby electronic equip-
ment may be affected.
CISPR B
Emissions Classification
Class B The System is suitable for use in all establishments including domestic
establishments.
Electromagnetic Immunity
The System is intended for use in the electromagnetic environment specified below. You should observe these requirements in the
use of the System.
Immunity against
IEC 60601-1-2
test level
Compliance level
(of this device) Electromagnetic environment
ElectroStatic Discharge,
ESD (IEC 61000-4-2)
contact discharge:
± 6 kV
air discharge: ± 8 kV
± 8 kV
± 15 kV
If floors are covered with synthetic material, try to avoid electro-
static discharges.
Power Frequency mag-
netic fields 50/60 Hz
(IEC 61000-4-8)
3 A/m 400 A/m Suitable for most environments. Magnetic field strengths in
excess of 400 A/m would be unlikely except in close proximity
to industrial magnetic devices.