User Manual
Table Of Contents
- Your New OmniPod Insulin Management System
- Getting Started
- The OmniPod Starter Kit
- Set Up the Personal Diabetes Manager (PDM)
- The Setup Wizard
- Enter your PDM ID
- Select the ID screen color
- Set date and time
- Enter basal settings
- Enter blood glucose sound setting and BG goal
- Set the suggested bolus calculator
- Enter target blood glucose value
- Enter minimum BG allowed for use in bolus calculation
- Enter insulin-to-carbohydrate ratio
- Enter correction factor
- Set reverse correction
- Enter the duration of insulin action
- Select bolus increment and enter maximum bolus
- Set extended bolus doses
- Set low reservoir advisory
- Set expiration notification
- Understanding and Adjusting Basal Rates
- Understanding and Delivering Bolus Doses
- Using the Personal Diabetes Manager
- Checking Your Blood Glucose
- The Built-in FreeStyle® Blood Glucose Meter
- The FreeStyle® Blood Glucose Test Strips
- The FreeStyle® Control Solution
- Performing a Control Solution Test
- Performing a Blood Glucose Reading
- Blood Glucose Results and the Suggested Bolus Calculator
- Entering Blood Glucose Readings Manually
- Editing Tags
- Low and High Blood Glucose Readings
- Important Health-Related Information
- Understanding Your Records
- Living with Diabetes
- Alerts and Alarms
- Communication Failures
- Appendix
- Pod Care and Maintenance
- Personal Diabetes Manager Care and Maintenance
- Storage and Supplies
- Suggested Bolus Calculator Examples and Guidelines
- OmniPod System Options and Settings
- Pod Specifications
- Accuracy Test Results
- Personal Diabetes Manager Specifications
- Blood Glucose Meter Specifications
- OmniPod System Label Symbols
- Personal Diabetes Manager Icons
- OmniPod System Notice Concerning Interference
- Electromagnetic Compatibility
- Customer Bill of Rights
- Limited Warranty for the Personal Diabetes Manager
- HIPAA Privacy Notice
- Glossary
- Index
Appendix
147
■ Personal Diabetes Manager Specifications
Size: 2.5" wide x 4.5" long x 1.0" high (6.4cm x 11.4cm x 2.5cm)
Weight (with batteries): 135 grams (4.76 oz)
Screen: 1.4" wide x 1.9" long (3.6cm x 4.8cm); 2.4" diagonal
(6.1cm) LCD
Battery: Powered by 2 AAA alkaline batteries
Battery life: Approximately 3 weeks
Operating temperature range: 40° F to 104° F (4.4° C to 40° C)
Storage temperature range: -20.2° F to 140° F (-29° C to 60° C)
Operating relative humidity range: 10% to 90% non-
condensing
Storage relative humidity range: 10% to 90% non-condensing
Operating atmospheric pressure: 696 hPA to 1062 hPA
Storage atmospheric pressure: 703 hPA to 1062 hPA
Communication distance: At start up the PDM and pod should
be adjacent and touching, either in or out of tray to ensure
proper communication during priming. At least 60" (152.40 cm)
during normal operation.
Alarm type: Audible
Alert type: Audible and vibratory
Warranty: 4 Year Limited (PDM)
■ Blood Glucose Meter Specifications
Assay method: Coulometric electrochemical sensor
Calibration: Plasma equivalent
Hematocrit: 15% to 65%
Measurement units: mg/dL
Result range: 20 to 500 mg/dL
Sample: Whole blood, capillary
Sample size: 300 nanoliters (0.3 microliter)
Test time: Results as fast as 7 seconds