- Understanding and Adjusting Basal Rates
- Understanding and Delivering Bolus Doses
- Using the Personal Diabetes Manager
- Checking Your Blood Glucose
- The Built-in FreeStyle® Blood Glucose Meter
- The FreeStyle® Blood Glucose Test Strips
- The FreeStyle® Control Solution
- Performing a Control Solution Test
- Performing a Blood Glucose Reading
- Blood Glucose Results and the Suggested Bolus Calculator
- Entering Blood Glucose Readings Manually
- Editing Tags
- Low and High Blood Glucose Readings
- Important Health-Related Information
- Understanding Your Records
- Living with Diabetes
- Alerts and Alarms
- Communication Failures
- Appendix
- Pod Care and Maintenance
- Personal Diabetes Manager Care and Maintenance
- Storage and Supplies
- Suggested Bolus Calculator Examples and Guidelines
- OmniPod System Options and Settings
- Pod Specifications
- Accuracy Test Results
- Personal Diabetes Manager Specifications
- Blood Glucose Meter Specifications
- OmniPod System Label Symbols
- Personal Diabetes Manager Icons
- OmniPod System Notice Concerning Interference
- Electromagnetic Compatibility
- Customer Bill of Rights
- Limited Warranty for the Personal Diabetes Manager
- HIPAA Privacy Notice
- Glossary
- Index
Alerts and Alarms
125
10
Auto-off.
Rem
ove Pod now.
Hazard
Alarm
Pod
and
PDM
Continuous tone. Pod has been deactivated. Insulin deliv-
ery has stopped. Press OK to acknowl-
edge.
Change Pod.
Check blood glucose.
Low reservoir.
Replace Pod soon.
(This is based on the vol-
ume level you selected.)
Advisory
Alarm
Pod
and
PDM
2 sets of beeps every minute for 3
minutes.
Repeats every 60 minutes until
acknowledged.
Escalates to Empty reservoir
hazard alarm when empty.
Press OK to acknowledge.
Empty reservoir.
Insulin delivery stopped.
Change Pod now.
Hazard
Alarm
Pod
and
PDM
Continuous tone. This alarm occurs when the Pod's
reservoir is out of insulin. Press OK to
acknowledge.
Change Pod.
Check blood glucose.
Alarm Type Source Beep Pattern Action
The Low reservoir alert will
escalate to an Empty reservoir
hazard alarm when insulin is
depleted. Be sure to respond to
the alert when it first occurs.