Checking Your Blood Glucose
80
7
Do not add water or any liquid to control solution.
Control solution tests are specified to be accurate only when
performed between 59° and 104° F (15° and 40° C).
The control solution test results should fall within the range
printed on the test strip vial label.
If control solution results are out of this range, REPEAT THE
TEST.
Out-of-range control solution results
Out-of-range test results may be caused by:
Expired or bad control solution
Expired or bad test strip
Error in performing test
Watered-down control solution
Code on test strip vial does not match code set in the PDM
Malfunction of the OmniPod System
Control solution test done outside 59º to 104ºF (15º to 40ºC)
If your control solution test results continue to fall outside the
range printed on the test strip vial:
1. The OmniPod System may not be working properly.
2. Do not use the System to test your blood glucose.
3. Call Customer Support, 800-591-3455, 24 hours/7 days
(from outside the United States, 781-457-5098).
Use only FreeStyle® control solution with the
OmniPod System. Other brands of control solu-
tion may produce inaccurate results with this
System.
Do not use control solution past the expiration
date or you may get inaccurate results.
Results from FreeStyle® control solution tests do
not reflect your blood glucose level.
The control solution range is a target range for control
solution only. It is not a target range for your blood
glucose.