59
CHAPTER 6
Using the Personal Diabetes Manager
The Status Screens
When a Pod is active and the PDM is turned on, the PDM estab-
lishes communication with the Pod to obtain a status check.
During the status check, the PDM collects information from the
Pod about bolus deliveries, active basal programs, and Pod expi-
ration. The information obtained is then displayed in the status
screen, which shows the Systems current operating status.
The Status screen (Figure 6-1) displays:
Time, date, and result of last blood glucose checked or
entered manually
Time, date, and total amount of last bolus delivered
Amount of insulin on board, if suggested bolus calculator is in
use
Name and rate of active basal program or temp basal
program
“Ext bolus,” the amount of insulin, and the delivery time
remaining if an extended bolus is being delivered
“INSULIN SUSPENDED” if insulin delivery has been
suspended
Time and date when the Pod will expire