Errors, Advisories, and Hazard Alarms
127
10
End of insulin suspend.
Press “Ok” to resume basal
rate delivery.
Pod 2 sets of beeps every minute for 3 minutes.
Repeats every 15 minutes until insulin delivery is
resumed.
Press OK to resume insulin delivery at
programmed basal rate.
Auto-off advisory Pod
and
PDM
2 sets of beeps every minute for 15 minutes.
Escalates to Auto-off hazard alarm if not
acknowledged within 15 minutes.
Press OK to acknowledge.
Reset clock.
Stuck key detected.
Check Home button to
resolve.
For technical support, call
800.591.3455.
PDM 2 sets of beeps every 5 minutes. Check the key described on the
screen. (In this example, the stuck key
is the Home button.) Press the key to
unstick it.
If this does not work, call Customer
Support at 800-591-3455 for assis-
tance (from outside the United States,
781-457-5098).
Alarm Source Beep Pattern Action