User's Manual

Set Diagnostic Functions
The Diagnostics screen allows you to instantly confirm how the
system is working or to completely update settings entered
during setup (see Chapter 2, Getting Started). They include:
Check alar
ms: Confirms that all alarms work properly when
needed. When you select this function, the remote beeps, then
vibrates, and then the pump beeps.
If the remote fails to beep and/or vibrate, call Insulet
immediately at (800) 591-3455. If a pump is active and fails
to beep, change pump immediately (see Chapter 5, The
iXL Pump).
Reset r
emote (soft reset): Restores all settings in the remote to
the factory defaults.
Resetting the remote deletes all history records, basal
programs, temporary basal presets, and user information
from memory. Before using this feature, be sure you have
a written record of the information you need before it is
deleted.
As a safety feature, you cannot reset the remote from the
Diagnostics screen when a pump is active. You must first
deactivate the pump.
Set diagnostic options
1. In the Main menu, use the Up/Down Controller button to
choose Settings; then press the soft key labeled Select.
2. Choose System setup, then press Select.
3. Choose Diagnostics, then press Select (Figure 6-8).
To check alarms:
1. Choose Check alarms, then press Select.
2. Press Confirm. The remote beeps three times, then vibrates.
If a pump is active, it then beeps three times.
Check the alarm function at every pump change.
56 iXL Diabetes Management System User Guide: IXL Remote Controller Features
iXL Remote Controller Features6