User's Manual

Change Bolus and Basal Settings
You can reset the bolus increment, the maximum bolus dose, the
maximum basal rate, and the maximum daily dose.
1. In the Main menu, use the Up/Down Controller button to
choose Settings; then press the soft key labeled Select.
2. Choose System setup, then press Select.
3. Choose Bolus/basal setup, then press Select.
4. Choose one of the following (Figure 6-4) and then press Select:
Bolus increment
Max bolus
Max basal rate
Max daily dose
5. Use the Up/Down Controller button to enter a value, then
press Select.
Some settings have preset defaults, but all settings
are changeable. See the Appendices for a list of system
specifications, including preset default settings.
6. Repeat steps 4 and 5 for each setting you want to change.
For details on basal rate and bolus dose settings, see
Chapter 3, Basal Rates, and Chapter 4, Bolus Doses.
If the remotes clock becomes corrupted and the clock
is reset, then the max daily dose checks will only apply
moving forward. If a clock reset is required due to a
Remote error, the max daily dose will be reset and the
max daily dose checks will only apply to insulin deliveries
moving forward from this point in time (see Chapter 9,
Hazard and Advisory Alarms).
52 iXL Diabetes Management System User Guide: IXL Remote Controller Features
iXL Remote Controller Features6
Figure 6-4