At the end of the suspend period, a pump Advisory alarm will
occur.
To resume your programmed basal rate, press OK (Figure 5-19).
The Advisory alarm will repeat every 15 minutes until you
press OK.
Resume Insulin Delivery
1. In the Main menu, use the Up/Down Controller button to
choose Resume (Figure 5-20); then press the soft key
labeled Select.
2. Press Resume to restart the basal program that was
previously running. Or press Back if you want to continue
the suspension.