7. Remove needle from vial and insert into the insulin fill port on
the underside of the pump body (Figure 5-10).
8. Depress syringe plunger to completely empty syringe into the
pump.
9. Remove needle from insulin fill port. Port is self-sealing;
insulin will not leak after needle is removed.
Do NOT insert the fill syringe into the fill port more than
once.
10.Place protective cap on needle and remove needle from
syringe.
11.Place capped needle in sharps container. Dispose of used
needles according to local waste disposal regulations.
The fill syringe is intended for single use only and should
be used only with iXL Pumps.
12.After you fill the pump, press Continue (Figure 5-11). The
pumps RediSet system performs a series of safety checks
and automatically primes the pump. Once complete, the
remote beeps, letting you know that the priming and safety
checks were successful.
40 iXL Diabetes Management System User Guide: The iXL Pump
The iXL Pump5
Figure 5-10
Figure 5-11