User's Manual
Basal Rates
Cancel a one-time temporary basal rate
1. In the Main menu, use the Up/Down Controller button to
choose Suspend/cancel; then press the soft key labeled
Select.
2. Choose Cancel temp basal (Figure 3-17); then press Select.
3. Press Confirm to cancel the temporary basal shown on the
screen.
Or press Back to back out of this process.
Maximum Basal Rate
What is the maximum basal rate?
The maximum basal rate is a safety feature. It limits the basal
rate (U/hr) that the iXL Pump can deliver. Once it is entered into
the remote’s memory, the maximum applies to both regular basal
programs and temporary basal rates.
You and your healthcare provider will enter an initial maximum
basal rate into your remote during the setup process (see
Chapter 2, Getting Started). You can change it later as you
fine-tune your system settings or when your needs change.
Set a maximum basal rate
1. In the Main menu, use the Up/Down Controller button to
choose Settings; then press the soft key labeled Select.
2. Choose System setup, then press Select.
3. Choose Bolus/basal setup, then press Select.
4. Choose Max basal rate, then press Select.
5. Enter a new maximum basal rate, then press Enter.
6. You can press Back at any screen to back out of this process.
iXL™ Diabetes Management System User Guide: Basal Rates 25
3
Figure 3-17