User's Manual
Be Aware: Hazard and Advisory Alarms
iXL™ Diabetes Management System User Guide: Be Aware: Hazard and Advisory Alarms 91
9
Advisory Alarms
Alarm Type Source Beep pattern and escalation process Action to resolve
Stuck key detected
Remote 3 sets of beeps every 3 minutes for 60 Check key to resolve.
minutes. If key remains stuck, press
Fail.
Escalates to “Stuck key” hazard alarm if
not resolved within 60 minutes of if you
press
Fail.
Respond to advisory alarms
For an advisory alarm caused by a low reservoir or pump
expiration
:
1. Acknowledge the alarm condition by pressing
OK,
which silences the alarm.
2. Deactivate and remove the current pump (see Chapter
5, The iXL Pump).
3. Attach and activate a new pump (see Chapter 5, The iXL
Pump).
For an advisory alarm caused by
Auto off or End of
suspend
, simply press OK.
For an advisory alarm caused by a
stuck key, press the
soft key or button named in the on-screen message. If
this does not resolve the problem, press
Fail. This action
results in a hazard alarm with instructions to call Insulet
support for assistance.