User's Manual
Advisory Alarms
90 iXL™ Diabetes Management System User Guide: Be Aware: Hazard and Advisory Alarms
Be Aware: Hazard and Advisory Alarms9
Alarm Type Source Beep pattern and escalation process Action to resolve
End of suspend
Pump 3 sets of beeps every minute for 3 minutes. Press OK to resume insulin
Repeats until insulin delivery is resumed. delivery at programmed
basal rate.
Auto off advisory Pump and 3 sets of beeps every minute for 15 minutes. Press OK to acknowledge.
Remote Escalates to “Auto off” hazard alarm if not
acknowledged within 15 minutes.