Alarm Type Source Beep pattern and escalation process Action to resolve
Low reservoir volume
Pump 3 sets of beeps every minute for 3 minutes. Press OK to acknowledge.
Repeats every 60 minutes until
acknowledged.
Escalates to hazard alarm when is empty.
(based on volume selected by user)
9
88 iXL™ Diabetes Management System User Guide: Be Aware: Hazard and Advisory Alarms
Advisory Alarms
Advisory alarms tell you about a condition that requires
your attention. When you hear an advisory alarm, turn on
the remote to display the
Status screen. The screen
displays a message describing the alarm and how to
resolve it.
Some advisory alarms are controlled by levels you set in
the
Settings screen (for example, the insulin level at which
the low-reservoir-volume alarm sounds). All advisory
alarms are time-sensitive and require a response. Some
advisory alarms escalate to hazard alarms if you do not
respond within a certain time period (as noted in the
table below).
Each advisory alarm beeps intermittently and displays an
on-screen message once the remote
Status screen is
updated.
Be Aware: Hazard and Advisory Alarms9
Advisory Alarms