Environmental and power specifications
Environmental conditions
Table 2.2: Storage Conditions
Power requirements
Compressed Air Requirements
Description Value
Operating altitude up to 2000 meters above sea level
Operating temperature range 16° to 30° C
Operating humidity Max 80% at 30°C
Overvoltage Category I
Pollution Degree 2
Description Value
Non Operating altitude up to 12,000 meters above sea level
Non Operating Humidity 5% to 90%
Non Operating Temperature -20°C to +60°C
Description Value
Mains Line Frequency 50/60 Hz
Mains Line voltages fluctuation Up to ± 10 %
• Mains Line Voltages depending on system configuration as follow:
Single phase configuration with UPS 230 V ~ 3680 VA
Single phase configuration without UPS 230 V ~ 9200 VA
Three phases configuration without UPS 400 V 3N ~ 27600
Mains Line voltages fluctuations Up to ± 10 %
Description Specification
Compressed air must be clean Oil free and dry
Compressed air pressure range 7 to 9 bar