Installation and Operation Manual
3.30.22 357-00060-01 Rev A © Inovonics, 2022 - www.inovonics.com 2
• The red LED will flash and the pendant will vibrate, indicating that an
alarm has been sent. (For more information, see section 3.1, “Send an
Alarm” on page 2.)
5. Clear the alarm sent when you activated the pendant per section 3.2,
“Clear an Alarm”.
2.3 Register the Transmitter
The transmitter sends a check-in message to the receiver or gateway every
60 minutes for the purposes of supervision. The transmitter must be
registered and supervised.
Caution: The EN2221S-60 location pendant must have a supervision
window of no more than 12 hours to avoid not discovering a missing device
within 24 hours.
Each transmitter has a unique factory-programmed identification number.
Refer to the receiver or gateway's installation instructions for details on
registering and setting up supervision.
1. When prompted by the receiver or gateway to reset transmitter, press
the activation button, and then clear the alarm per section 3.2, “Clear an
Alarm”.
Caution: The transmitter should be tested after registration to ensure
operation. See section 6, “Test the Transmitter”.
2.4 Wear the Transmitter
The pendant can be worn in three ways:
• Around the neck with the neck lanyard, which is included with the
location pendant with neck lanyard.
Caution: The neck lanyard included with the pendant is designed with a
breakaway feature for user safety. Substitution of a stronger cord or chain
may result in injury to the wearer.
• On the arm with the wristband, which comes attached on the pendant
with wristband.
• On the belt, using the optional ACC683 belt loop attachment kit.
Neck Lanyard
To attach the neck lanyard:
2. Secure the lanyard to the pendant’s attachment loop with a simple girth
hitch knot.
Figure 4 A simple girth hitch knot
3. Adjust the neck lanyard to the wearer’s neck size.
• Pull both slide adjustment tabs on the lanyard to reduce the length of
the cord.
• Pull both sides of the lanyard cord to increase the length of the cord.
Figure 5 Adjust the neck lanyard
Wristband
The wristband comes attached to the location pendant with wristband. The
wristband is a standard 16mm width, allowing for replacement with a
wristband of the user’s choice. The standard length wristband is designed
to fit wrist circumferences from 5.5 to 7.5 inches; the long wristband is
designed to fit wrist circumferences from 6.5 to 9.25 inches.
Note: Replacement wristbands must be of a similar thickness to the
provided 16mm wristband. Metal expansion bands will not attach reliably to
the device.
3 Operate the Transmitter
3.1 Send an Alarm
To send an alarm:
1. Press the activation button for at least one second.
• When activated, the pendant will vibrate briefly and the red
transmission LED will blink rapidly for the first five seconds, and then
slowly until the alarm is cleared.
3.2 Clear an Alarm
Note: After sending an alarm, an end user can send a subsequent alarm
once ten seconds have elapsed, even if the initial alarm has not been
cleared on the device.
There are two ways to clear an alarm:
Using the Alarm Clearance Card
Note: The ACC680 alarm clearance card is an optional accessory sold
separately.
1. Place the dot on the ACC680 alarm clearance card over the activation
button.
• The red transmission LED will cease flashing and the blue alarm clear
LED will flash quickly six times to indicate the alarm has cleared.
Figure 6 Place the alarm clearance card over the activation button
Using a Button Press Pattern
1. Press the activation button three times, quickly.
2. When the blue LED flashes twice, press the activation button three
more times, quickly.
• The blue alarm clear LED will flash quickly six times to indicate the
alarm has cleared.
3.3 Coin Cell Low Battery Alert
When a low battery is detected, the transmitter will send a low battery alert,
and start a seven day countdown. If the battery has not been replaced
within two days from when the low battery alert message was sent, the
yellow low battery LED on the bottom of the pendant will begin to blink, and
will not stop until the battery is replaced, or the seven day countdown
expires.
If the seven day countdown expires without the battery being changed, a
high priority shutdown message will be sent to the application software and
the pendant will automatically go into storage mode. For more about
storage mode, see section 5, “Storage Mode”.
4 Coin Cell Battery Replacement
Note: Inovonics has tested and recommends Panasonic®, Maxell® and
FDK® (formerly Sanyo) CR2032 coin cell batteries.
The waterproof pendant uses one standard CR2032 coin cell battery. To
change the battery:
1. Use the ACC680 alarm clearance card or a quarter to turn the battery
door to the unlocked padlock icon.
2. Remove the battery door.
3. Remove the old battery from the battery compartment.
4. Place the new battery in the battery compartment, ensuring that the
positive terminal (+) faces up.
5. Seat the battery door over the battery so that the arrow on the battery
door is lined up with the unlocked padlock icon.
Cord
Attachment
loop
Tighten
Loosen