Product Manual

14
NORMAL OPERATION INDICATORS
During normal operation, the Control Panel and Display Screen
should appear as described below:
OPERATION DESCRIPTION
Active or Rest Mode
Activitated
The word “Active Mode” or “Rest Mode” will appear next to your selected
pulse setting. If “Active Mode” is displayed it means your breath rate is
typical for an active user. If “Rest Mode” is displayed it means your breath
rate is slower than your normal active breath rate.
Alarm Muted
The audible alarm signal will be muted and the Mute symbol will
display on screen. The red alarm LED will remain illuminated on the
Control Panel and the alarm title will remain on the display.
Breath Detected
The PULSE-WAVE Breath Detection Symbol will flash as you breathe.
This is confirmation the device is delivering a dose of oxygen with each
inhaled breath.
If you stop breathing through the nasal cannula (mouth breathing),
the unit will begin delivering doses of oxygen automatically without
the PULSE-WAVE symbol. After 75 seconds, the unit will provide a
“NO BREATH DETECTED” alarm and will continue to automatically
provide doses of oxygen until normal breathing resumes. When the
unit detects normal breathing, the “NO BREATH DETECTED” alarm
will cease and the PULSE-WAVE symbol will indicate detection of your
breath.
Charging Status
The following messages will appear on the display depending on the
battery being used or charged.
“Charging INT” means the Internal Battery is charging.
“Charging EXT” means the External Battery is charging.
“Running INT” means the Internal Battery is in use.
“Running EXT” means the External Battery is in use.
NOTE: By design, the device will preserve the Internal Battery charge as
long as possible. This means that the External Battery will deplete before
the Internal Battery and the Internal Battery will charge before the External
Battery when connected.
Power Interruption
The device will emit a two second audible tone if the AC or DC Power is
interrupted or intentionally unplugged. The unit will automatically switch
to battery operation. If the power was not intentionally removed, check
all connections between the unit and the power source to ensure all are
secure and firmly plugged in. If the power indicator icon does not show
on the display, switch to another power source (AC or DC) or source of
oxygen and contact your equipment provider.
Powered On The Power Indicator is illuminated green and the Display Screen is on.
Setting Selected “1”, “2”, “3” or “4” will appear under “Setting” on the Display Screen.