Product Manual
2. AC Power Supply: This power supply allows the device to be powered
and charged from a standard 100-240 VAC,
50/60 Hz electrical outlet. When using this
option, power from the AC outlet powers
the unit and recharges the Internal and/
or External Battery. The External Power
symbol will also appear on the display
screen when connected to this power
supply.
NOTE: Using an AC Power Supply that is not specified by Inova Labs
may damage the device and will void all warranties.
3. DC Power Supply: This power supply can be connected from the device
to a vehicle’s (car, boat, motor home, etc.)
standard 12 VDC outlet. When the system
is connected to this option, power from the
vehicle battery powers the device and
recharges the Internal and/or External
Battery. The External Power symbol will
also appear on the display screen when
connected to this power supply.
NOTE: The 12 VDC outlets in some vehicles will not provide sufficient
current to operate the device and charge the battery at the same time.
Verify that a DC outlet can provide at least 10 amps at 12 VDC. Car
auxiliary power varies significantly; therefore, the rate of charging will
vary. It is recommended that the vehicle be turned on before plugging
the DC Power Supply into the device.
4. External Battery: This power supply connects directly into the device’s
External Battery Port. When the system is connected
to this option, the display will read “Running EXT”
and the EXT Battery Bar will appear. A fully charged
External Battery can provide up to 4.75 additional
hours of power to the device depending on breath
rate.
The External Battery is maintenance-free
and replacements can be purchased
from your equipment provider.
10
OPERATING
INSTRUCTIONS
continued
4