Product Manual

9
OPERATING INSTRUCTIONS
The LifeChoice Activox POC must be operated and stored in its carry case
at all times.
BEFORE FIRST USE:
1. Fully charge the device. It can take up to 4 hours to reach a full
charge on the Internal Battery. Begin charging by connecting
the AC Power Supply to the Charger Port on the device and an
electrical outlet.
2. Verify that the INT Battery Bar, the External Power Symbol and the
message “Charging INT” appear on the display. This indicates that
the system recognizes the external power source and is charging the
Internal Battery.
If an External Battery was purchased, connect it to the External
Battery Port. By design, the External Battery will begin charging after
the Internal Battery has reached a full charge. When connected, the
External Battery Bar, the External Power Symbol and the message
“Charging EXT” appear on the display. It can take up to 2 hours to
fully charge the External Battery.
CAUTION: Never use an external battery that is not authorized by
Inova Labs Inc.
POWERING YOUR POC
The device can be powered from four different sources: (1) Internal
Battery, (2) AC Power Supply, (3) DC Power Supply, and (4) External
Battery (optional, sold separately from the Standard Package).
1. Internal Battery: A rechargeable internal battery is located within
each device. When the device is being powered from the Internal
Battery, the display will read “Running INT” and the INT Battery
Bar will appear.
A fully charged Internal Battery
can provide up to 10.25 hours of
runtime depending on breath rate.
(Reference Battery Life Timetable
on page 12.)
If not connected to the AC or DC Power Supply, the Internal
Battery will slowly discharge over time even when not in use.
Always check the battery level prior to use to ensure adequate charge
level.
The Internal Battery is maintenance-free and can only be replaced at
the factory or by an authorized repair facility.
4